DBD (Dielectric Barrier Discharge) cold plasma therapy devices require careful parameter configuration and proper operating technique to achieve optimal treatment results and ensure a comfortable experience for the patient. Based on DGE product technical manuals, this guide systematically presents gas flow parameter settings, recommended treatment durations by wound type, standard operating procedures, and safety precautions for veterinary clinical reference.

1. Gas Flow Parameters

The therapy device uses helium as its working gas. The combination of helium regulator pressure and internal flow rate setting determines plasma jet intensity and coverage area. The table below is based on a 10L helium tank:

Helium Regulator Pressure (MPa) Internal Flow Setting Helium Flow Rate (L/min) Usage Duration (minutes)
0.1 Level 1 6 245
0.1 Level 2 8 185
0.2 Level 1 10 147
0.2 Level 2 13 114
⭐ Recommended Setting: Helium regulator at 0.1 MPa, internal flow Level 1, intensity Level 1. This combination keeps helium consumption low while delivering effective treatment — a single 10L tank supports approximately 245 minutes of therapy.

Parameter Selection Principles

2. Recommended Treatment Duration

Different wound types and sizes require different treatment durations. The following table provides clinically recommended reference values:

Wound Type Classification Recommended Duration
Sutured Wounds Up to 5 cm in length 30 seconds
5–15 cm in length 1–2 minutes
Over 15 cm in length 2–3 minutes
Open Wounds Small (approx. 1×1 cm) 30 seconds
Medium (approx. 3×3 cm) 1–2 minutes
Large (approx. 5×5 cm) 3–4 minutes

3. Operating Guidelines

The following methods apply to most clinical scenarios and are recommended wherever conditions permit.

Standard Operating Procedure

  1. Press the foot pedal to activate the device and observe the helium plasma jet emitted from the treatment handpiece.
  2. Bring the plasma jet into contact with the skin. A bright spot should be visible on the skin surface.
  3. Key Point: The optimal treatment distance is achieved when the bright spot is located at approximately one-third of the visible plasma jet length. This contact point is called the treatment center point.
  4. The treatment center point must be kept in continuous motion to avoid prolonged local exposure that could cause a burning sensation.
  5. Maintain this distance and gently sweep the treatment center point across the affected area.
  6. Repeating the sweep 2–3 times completes one treatment session.

Suitable Wound Types

Treating Large Irregular Wounds

For large, irregularly-shaped open wounds, use a zone-by-zone sweeping approach. Sequentially cover different areas of the wound with the treatment center point, ensuring uniform exposure across the entire wound surface while avoiding missed areas or over-treatment.

4. Contraindications & Safety

⚠️ Absolute Contraindication: Avoid treating areas with conductive metal implants or other conductive materials.
Do not direct the cold plasma jet directly onto any metal implant.

Precautions

Disclaimer: The parameters in this guide are for reference only. Specific treatment plans should be determined by a licensed veterinarian based on the individual patient's condition. Please read the product manual carefully before use.